Sign in or register

My Eezglobal

My shopping cart

   
Home&Garden,Lab > Lab Equipment> Test, Measurement & Inspection> Impact Testing Machine>
022537:Analog Impacting Testing Machine
Unit:
set
Price:
1,169.00$
Weight:
48.00kg
QTY:
Out Of Stock
     

Similar products

056409:110V 60Hz drop impact tester for ball impact testing machine

056491:Pipe impact testing machine

20140320

Analog Impacting Testing Machine

 


 Application and scope of application:

It is determined resisting impact strength of the engineering plastics, the glass fiber, the ceramics, the cast stone, the plastic electric appliance, the insulation material, plastic pipes and nonmetallic material etc. The products according with standards requirements of ISO 180 / GBT 1843 / GBT2611 / JBT8761 etc.

Specification

Project

Specification

capacity

66lb(30kg110lb(50kg),330lb(150 kg)-cm

Impact velocity:

3.4 m/sec(11ft/sec)

Impact end to axis distance

12.87’’(32.7 cm)

Angle

150°

Volume

19.7*12.6*29.5inch(50 cm ×32 cm ×75 cm)

Edge radius

R=0.8±0.2mm

Impact blade angle

37°±1°

Pendulum energy

2.75J ,5.5J,11J

Weight

126lb(57kg)

Use requirements:

 The impact of a hammer at an angle of 150 and an angle of less than an angle of150 at the other end.

The decrease reason is as follows

1.The sample itself has impact resistance (impact value);

2.The friction of the shock process, that is, the loss of wind resistance;

3.The pointer friction of the shock machine, that is, the inertia loss.

Therefore, the impact value of the sample can be calculated by the above factor

                                           Maintenance

1)、cleaningKeep the instrument clean and clean with cotton wiping instrument before and after each test.

2)、AntirustCheck the normal operation of the relative motion parts of the instrument and add the lubricating oil

                         Troubleshooting

situation

Possible cause

Troubleshooting

Test value is not allowed.

 

The instrument is not put on the level

The weight of the hammer is inaccurate.

The wobble performance of the hammer is bad.

The slide performance of the pointer is bad.

Adjust the instrument level of the base screw, put the instrument level.

Add lubricating oil or adjust weight.

Replace or repair the pointer.

Adjust the position of the specimen and fit the tightness.

The hammer can not be fixed on the fixed hook.

Spring elasticity of fixed hook.

Replace the fixed hook part spring.


Direction


 


Item Included:(022537)

  • Analog Impacting Testing Machine x1

     